Specifications

 

Click on the ‘+’ symbol within each section to review the IDS-iSYS System Specifications

  • Random access
  • Time to first result: < 25 minutes (assay dependent)
  • Up to 120 tests/hour (assay dependent)
  • > 600 results per day (assay dependent)
  • Wavelengths: from 300 to 500 nm
  • Linearity: Up to 10 million RLU
  • Spectrophotometer with LED technology
  • 12 wavelengths interferential filters: 340, 405, 450, 500, 540, 550, 580, 620, 650, 700, 720 and 750 nm.
  • Linearity: Up to 3 Abs
  • Bi-chromatic mode availability
  • Light sources: multiple LEDs
  • Serum, plasma, urine or saliva (assay dependent)
  • Sampling on primary tubes, secondary tubes, micro cups, paediatric tubes (assay dependent)
  • Default sample tube library: Primary Tubes: 4.5 mL, 6 mL, 8.5 mL, 10 mL
  • Secondary Tube: 5 mL
  • Micro cups: 500 μL, 2 mL
  • Further sample tubes can be easily configured
  • 120 positions for samples, calibrators and controls (6 racks with 20 positions)
  • Possibility of STAT sample at any position
  • Integrated reader in sample compartment for automatic reading of barcoded sample tubes (positive identification)
  • Barcode types: Code 2/5, EAN/UPC, Code 39, Code 128 and Codabar
  • On-board dilution and automatic pre-treatment
  • Up to 15 different immunoassay cartridges
  • Integrated reader for automatic reading of reagent cartridge barcodes
  • On-board storage: 12 – 15 °C during operation, 8 – 10 °C in standby mode (configuration dependent)
  • Single pipetting arm for samples and reagents
  • Liquid level detection by capacitance (samples and reagents)
  • Clot detection (by pressure transducer)
  • Internal and external rinsing between each pipetting of sample
  • Possibility of specific rinses using D-SORB solution
  • Preheating of sampling (reagents & samples)
  • Sample volumes: from 4 μL to 300 μL (assay dependent)
  • Temperature regulated at 37°C ± 0.5°C
  • 90 positions for disposable cuvettes
  • Automatic cuvette supply by cuvette loader
  • 4 independent washers
  • Loader for 1 x cube of cuvettes
  • 1 x cube contains 960 disposable cuvettes
  • 960 cuvettes on board
  • Preheated loader
  • IDS-iSYS System Liquid (Syst L)
  • IDS-iSYS Wash Solution (Wash S)
  • IDS-iSYS Trigger Set
  • IDS-iSYS Cartridge Check System (CCS)
  • D-SORB Solution
  • Reader on system front face for barcode identification, cuvettes, and ancillary reagents
  • 10 litre container for liquid waste
  • Solid waste (cuvettes) disposed of in solid waste drawer
  • User level management
  • Continuous access to programming
  • Full traceability: patient results, calibrations, controls.
  • Automatic calibration. QC management: up to 4 levels, analysis by Levey-Jennings and Westgard rules (configurable)
  • Lot management of reagents, calibrators and controls by barcodes
  • Reagent stability management
  • Local storage of 5000 patient results (including traceability)
  • Connection to LIS: bi-directional communication (ASTM protocol)
  • Windows XP Pro Service Pack 2
  • Windows Vista Service Pack 1
  • Windows 7
  • Windows 10
  • System: L 105 x H 70 x W 75 cm
  • Weight: 103 kg
  • PC: L 60 x H 40 x W 50 cm
  • Voltage: 100 – 240 V
  • Frequency: 50 – 60 Hertz
  • Maximum power consumed: 750 VA